Test Code LAB318 PLATELET FUNCTION ANALYSIS
Additional Codes
Orderable Codes:
| Epic Code | Atlas Code | Mayo Access ID | Order LOINC |
| LAB318 | PFA1 | FAH5425 | 72257-9 |
Result Component(s):
| Reporting Name | Epic Code | Atlas Code | Mayo Access ID | Result LOINC |
| COL/EPI Cartridge | 12301002503 | PFAEPI | FAH5426 | 24472-3 |
Specimen Information
| Container | Specimen | Temperature | Collect Vol | Submit Vol | Min Vol | Stability |
| Blue Top Tube | Whole Blood | Ambient | To fill line | To fill line | To fill line | 4 hours |
Testing must be performed within 4 hours of sample collection. Requires a separate aliquot if other coagulation tests are requested.
Platelet Function Analysis (PFA) is strongly dependent on the correct method of blood collection requiring venipuncture through a19-21G needle drawn directly into a full size evacuated sodium citrate tube. Do not draw through a line or port, which may cause platelet clumping and result in specimen rejection. Ensure proper specimen mixing by gently inverting by hand 4 times. Must remain unspun at ambient temp. Testing must be performed within 4 hours of sample collection. PFA requires separate tube; additional samples must be collected if other coagulation tests are requested.
Test Schedule / Analytical Time / Test Priority
Daily / 24 Hours / Available STAT
Method
PFA-100
Instrumentation
PFA-100
Reference Range
COL/EPI: 94 – 193 Seconds
COL/ADP: 71 – 118 Seconds
Section
Coagulation
Performing Location
University of Vermont Medical Center
Is the UVMMC lab NY State Certified to perform this testing? Yes/No
Yes
CPT Code(s)
| Description | CPT Code |
| Platelet Function Analysis | 85576 |