Sign in →

T3 (Triiodothyronine), Reverse, Serum

Additional Codes

Primary ID

Epic Code

Mayo Test ID

T3REV

LAB138 RT3

 

Reporting Name

T3 (Triiodothyronine), Reverse, S

Useful For

Aiding in the diagnosis of the sick euthyroid syndrome

Method Name

Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)

Portions of this test are covered by patents held by Quest Diagnostics.

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Specimen Required


Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube:

Preferred: Red top

Acceptable: Serum gel

Submission Container/Tube: Plastic vial

Specimen Volume: 0.8 mL

Collection Instructions: Centrifuge and aliquot serum into a plastic vial.


Specimen Minimum Volume

0.4 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 28 days
  Ambient  28 days
  Frozen  28 days

Reject Due To

Gross hemolysis Reject
Gross lipemia OK
Gross icterus OK

Reference Values

10-24 ng/dL

Day(s) Performed

Monday through Friday

CPT Code Information

84482

LOINC Code Information

Test ID Test Order Name Order LOINC Value
RT3 T3 (Triiodothyronine), Reverse, S 3052-8

 

Result ID Test Result Name Result LOINC Value
9405 T3(Triiodothyronine),Reverse,S 3052-8

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

Report Available

2 to 6 days

Specimen Retention Time

14 days

NY State Approved

Yes

Forms

If not ordering electronically, complete, print, and send a General Request (T239) with the specimen.